Corrected Intraocular Pressure Variability with Central Corneal Thickness Measurement

利用中央角膜厚度测量校正眼内压变异性

阅读:1

Abstract

PURPOSE: To evaluate variability in measured intraocular pressure (IOP) values when correlated with central corneal thickness measurements obtained by both ultrasonic and optical tools. PATIENTS AND METHODS: We included 46 eyes of 46 healthy subjects (age range 35-76 years). Exclusion criteria for the current study were patients with confirmed diagnosis of glaucoma, glaucoma suspect patients and those having corneal opacities, scars or prior cornea-based laser vision correction. Central corneal thickness (CCT) was measured using two methods: ultrasonically (CCT1) by pachymeter and optically (CCT2) using anterior-segment optical coherence tomography (AS-OCT). The IOP was measured in all subjects using Goldmann applanation tonometry (GAT). RESULTS: No significant difference was detected comparing intraocular pressure values and CCT-corrected intraocular pressure (IOPcc1 and IOPcc2) (P=0.47 and P=0.06, respectively) among the study participants. A significant negative correlation was found between corneal thickness-corrected IOP values and the measured central corneal thickness by both optical and ultrasonic tools (P=0.004 and P=0.001, respectively). CONCLUSION: Intraocular pressure appears to be dependent and positively correlated with CCT changes. However, this does not appear to depend largely on methods used for measuring the CCT in the current study. Corrected intraocular pressure is negatively correlated to both ultrasonically and optically measured central corneal thickness.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。