A comparison of manifest refractions, cycloplegic refractions and retinoscopy on the RMA-3000 autorefractometer in children aged 3 to 15 years

对 3 至 15 岁儿童使用 RMA-3000 自动验光仪进行显性验光、睫状肌麻痹验光和检影验光的比较

阅读:1

Abstract

PURPOSE: The study was conducted to compare the accuracy of readings of the RMA-3000 autorefractometer (Topcon, Tokyo, Japan) with traditional retinoscopy as a means of determining the approximate subjective refraction in children after cycloplegia. METHODS: 142 children aged 3 to 15 years were included. All children had their refractive status measured with the RMA-3000 autorefractometer (noncycloplegic autorefraction [AR]). Subsequently all children underwent cycloplegia and the refractive status was estimated again with the autorefractometer (cycloplegic autorefraction [ARC]) and traditional retinoscopy (RC) by examiners who were unaware of the results from the other techniques. RESULTS: From 69 right eyes with negative sphere we observed that the sphere power was significantly higher (more than 0.5 diopters) in AR than in ARC (P = 0.0001) and RC (P = 0.0001). From the 73 normal and hyperopic right eyes we observed that the sphere power was significantly lower (more than 0.5 diopters) in AR than in ARC (P = 0.0001) and RC (P = 0.0001). CONCLUSIONS: The use of the autorefractometer in children (in whom accommodation is more active than older patients) without cycloplegia may underestimate the actual hyperopia and overestimate the actual myopia. Manual retinoscopy is still the most accurate technique to estimate refractive status in children.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。