Comparison of intraocular pressure measurements with the portable PT100 noncontact tonometer and goldmann applanation tonometry

便携式PT100非接触式眼压计与Goldmann压平式眼压计测量眼压的比较

阅读:1

Abstract

PURPOSE: Noncontact tonometers are useful when regulations preclude use of contact tonometers by medical students and other nonophthalmologists. Our study compared the measurements by the portable, noncontact tonometer (PT100) with Goldmann applanation tonometry (GAT). METHODS: This was a prospective study of 98 eyes from 98 patients. Intraocular pressure (IOP) was measured by GAT and the PT100 (Reichert, Buffalo, NY). RESULTS: Mean IOP measurements showed no significant differences in measurements performed by the two tonometers (P = 0.64). Measurements by the two tonometers were in agreement by

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。