Comparison of single- and multi-isocenter planning with Dynamic WaveArc for multiple brain metastases

动态 WaveArc 技术在多发性脑转移瘤治疗中的单等中心和多等中心规划比较

阅读:1

Abstract

Dynamic WaveArc (DWA) is a technique used for continuous, non-coplanar volumetric-modulated arc therapy on the Vero4DRT platform. This study aimed to evaluate the application of single-isocenter DWA (SI-DWA) for treating multiple brain metastases by comparing dose distribution and irradiation time with multi-isocenter DWA (MI-DWA) through retrospective treatment planning. Treatment plans were developed for SI-DWA and MI-DWA in 14 cases with 3-5 brain metastases. Parameters assessed included target dose indices, such as conformity index (CI) of the planning target volume (PTV), volumes of normal brain excluding gross tumor volumes (GTVs) receiving a single dose equivalent of 14 Gy (V14), V30%, V20%, V10%, volumes of normal brain, including GTVs receiving a single dose equivalent of 12 Gy (V12), D2% for other organs at risk, and beam-on time. SI-DWA showed inferior CI, V14, and V12 values for lesions with PTV volumes <1 cc, whereas it performed equivalently to MI-DWA for lesions with PTV volumes ≥1 cc. SI-DWA resulted in higher volumes of normal brain receiving low doses compared to MI-DWA. SI-DWA exhibited significantly shorter beam-on times than MI-DWA. In conclusion, SI-DWA is an effective method for treating multiple brain metastases with PTV volumes ≥1 cc, offering an index of radiation-induced brain necrosis comparable with MI-DWA while allowing for shorter irradiation times.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。