Evaluation of the detection accuracy of set-up for various treatment sites using surface-guided radiotherapy system, VOXELAN: a phantom study

利用体模引导放射治疗系统VOXELAN评估不同治疗部位的定位检测精度:一项体模研究

阅读:2

Abstract

The purpose of this study is to evaluate the detection accuracy of a 3-dimensional (3D) body scanner, VOXELAN, in surface-guided radiotherapy (SGRT) of each part of the human body using a whole-body human phantom. We used A Resusci Anne was used as the whole-body phantom. The detection accuracy of VOXELAN in a radiotherapy treatment room with a linear accelerator (LINAC) was evaluated for two reference images: reconstruction of the planning computed tomography (CT) image (CT reference) and scanning by VOXELAN before the treatment (scan reference). The accuracy of the translational and rotational directions was verified for four treatment sites (open face shell, breast, abdomen, and arm), using the magnitude of the 6D robotic couch movement as the true value. Our results showed that the detection accuracy improved as the displacement from the reference position decreased for all the sites. Using the scan reference, the average accuracy of the translational and rotational axes was within 1.44 mm and 0.41°, respectively, for all sites except the arms. Similarly, using the CT reference, the average accuracy was within 2.45 mm and 1.35°, respectively. Additionally, it was difficult for both reference images to recognize misalignment of the arms. In conclusion we discovered that VOXELAN achieved a high detection accuracy for the head with an open face shell, chest, and abdomen, indicating that the system is useful in a clinical setting. However, it is necessary to pay attention to location matching for areas with few features, such as surface irregularities and potential errors, when the reference image is created from CT.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。