Coexistence of R-Wave Oversensing and Undersensing in an Implantable Loop Recorder: The Issue of Multiple Sensing in Implantable Loop Recorders

植入式循环记录仪中R波过感知和欠感知共存:植入式循环记录仪中的多重感知问题

阅读:1

Abstract

BACKGROUND Implantable loop recorders (ILRs) are important tools in diagnosing unexplained syncope and palpitations through prolonged rhythm monitoring. However, ILRs can produce false-positive arrhythmia detections, due to signal oversensing and undersensing, which can complicate clinical interpretation. Oversensing typically involves P waves, T waves, or myopotentials; however, simultaneous oversensing and undersensing of multiple waveform components is rare. We report a case of R-wave oversensing of P waves and T waves, with undersensing of QRS complexes, resulting in erroneous atrial fibrillation (AF) alerts. CASE REPORT A 73-year-old woman with hypertension and hyperlipidemia presented with recurrent lightheadedness, palpitations, and hypotension. After an unrevealing initial workup, a Biotronik Biomonitor III ILR was implanted. Within 3 months, the device flagged over 200 episodes of AF and several ventricular fibrillation episodes, despite the patient remaining asymptomatic. Careful signal review revealed a triple-sensing issue of cardiac signals, leading to misclassification of arrhythmias. The sensing filter was reprogrammed from 10 Hz to 24 Hz, resulting in elimination of false-positive detections. Following reprogramming, the patient remained asymptomatic with no further inappropriate arrhythmia alerts and did not require additional interventions. CONCLUSIONS This case highlights a rare example of R-wave oversensing and undersensing by an ILR, emphasizing the importance of individualized device programming to optimize diagnostic accuracy. Additionally, it illustrates that unconventional implantation sites, such as the right parasternal region, can produce unique sensing challenges. Careful review of device signals and appropriate reprogramming can correct oversensing issues, preventing unnecessary interventions and improving patient management.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。