The mechanical property and corrosion resistance of Mg-Zn-Nd alloy fine wires in vitro and in vivo

镁锌钕合金细丝的体外和体内力学性能及耐腐蚀性能研究

阅读:1

Abstract

Titanium and its alloy are commonly used as surgical staples in the reconstruction of intestinal tract and stomach, however they cannot be absorbed in human body, which may cause a series of complications to influence further diagnosis. Magnesium and its alloy have great potential as surgical staples, because they can be degraded in human body and have good mechanical properties and biocompatibility. In this study, Mg-2Zn-0.5Nd (ZN20) alloy fine wires showed great potential as surgical staples. The ultimate tensile strength and elongation of ZN20 alloy fine wires were 248 MPa and 13%, respectively, which could be benefit for the deformation of the surgical staples from U-shape to B-shape. The bursting pressure of the wire was about 40 kPa, implying that it can supply sufficient mechanical support after anastomosis. Biochemical test and histological analysis illustrated good biocompatibility and biological safety of ZN20 alloy fine wire. The residual tensile stress formed on the outside of ZN20 fine wire during drawing would accelerate the corrosion. The second phase had a negative influence on corrosion property due to galvanic corrosion. The corrosion rate in vitro was faster than that in vivo due to the capsule formed on the surface of ZN20 alloy fine wire.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。