The performance of an inexpensive spark-induced breakdown spectroscopy instrument for near real-time analysis of toxic metal particles

一种低成本火花诱导击穿光谱仪在近实时分析有毒金属颗粒方面的性能

阅读:1

Abstract

To meet the demand for identifying and controlling toxic air contaminants in environmental justice communities, we have recently developed a cost-effective spark-induced breakdown spectroscopy (SIBS) instrument for detecting and quantifying toxic metal air pollutants. We characterized the detection limit and linearity of this SIBS instrument by analyzing nebulized elemental standard solutions. The experimental parameters affecting SIBS performance were optimized, including the time delay to observation, the distance between electrodes, and the ablation voltage. The instrument successfully detected Cr, Cu, Mn, Fe, Zn, Co, and Ni, with limits of detection ranged from 0.05 μg m(-3) to 0.81 μg m(-3) at a flow rate of 15 l min(-1) and a 30 min sampling duration. Similar to other investigations using ion breakdown spectroscopy, we did not observe strong emissions lines for As, Sb, Se, Hg, Pb, and Cd, which were likely due to spectral overlap, matrix effects, and the limited detection range of the optical components. Overall, SIBS is a promising technique for field measurements of toxic metals for environmental justice, industrial and urban applications.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。