Measurement of sodium ion concentration in undiluted urine with cation-selective polymeric membrane electrodes after the removal of interfering compounds

用阳离子选择性聚合物膜电极去除干扰物后测量未稀释尿液中的钠离子浓度

阅读:3
作者:Feyisayo Phillips, Kim Kaczor, Neel Gandhi, Bradford D Pendley, Robert K Danish, Michael R Neuman, Blanka Tóth, Viola Horváth, Ernö Lindner

Abstract

The measurement of sodium ion concentration in urine can provide diagnostic information and guide therapy. Unfortunately, neutral-carrier-based ion-selective electrodes show a large positive drift and loss in selectivity in undiluted urine. The extraction of electrically neutral lipids from the urine into the sensing membrane was suggested as the main source of the drift, loss of selectivity and the consequent incorrect concentration readings. In this work, (i) solvent-solvent extraction, (ii) membrane-immobilized solvent extraction and (iii) solid phase extraction were used to remove interfering compounds from urine samples. The "cleaned" urine samples were subsequently analyzed using a calixarene (sodium ionophore X)-based, solid-contact, sodium-selective electrode in a flow-through manifold. The solid-contact sodium sensors had excellent stability in cleaned urine and an acceptable bias compared to commercial clinical analyzers.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。