Monitoring Fetal Heart Rate during Labor: A Comparison of Three Methods

分娩过程中胎心率监测:三种方法的比较

阅读:1

Abstract

The purpose of the study was to compare the accuracy of a noninvasive fetal heart rate monitor with that of ultrasound, using a fetal scalp electrode as the gold standard, in laboring women of varying body habitus, throughout labor and delivery. Laboring women requiring fetal scalp electrode were monitored simultaneously with the investigational device (noninvasive fetal ECG), ultrasound, and fetal scalp electrode. An algorithm extracted the fetal heart rate from the noninvasive fetal ECG signal. Each noninvasive device recording was compared with fetal scalp electrode with regard to reliability by positive percent agreement and accuracy by root mean squared error. Seventy-one women were included in this analysis. Positive percent agreement was 83.4 ± 15.4% for noninvasive fetal ECG and 62.4 ± 26.7% for ultrasound. The root mean squared error compared with fetal scalp electrode-derived fetal heart rate was 4.8 ± 2.0 bpm for noninvasive fetal ECG and 14.3 ± 8.2 bpm for ultrasound. The superiority of noninvasive fetal ECG was maintained for stages 1 and 2 of labor and increases in body mass index. Compared with fetal scalp electrode-derived fetal heart rate, noninvasive fetal ECG is more accurate and reliable than ultrasound for intrapartum monitoring for stages 1 and 2 of labor and is less affected by increasing maternal body mass index. This confirms the results of other workers in this field.