Current practices and prospects for standardization of the hematopoietic colony-forming unit assay: a report by the cellular therapy team of the Biomedical Excellence for Safer Transfusion (BEST) Collaborative

造血集落形成单位测定标准化的当前做法和前景:生物医学卓越安全输血 (BEST) 合作组织细胞治疗团队的报告

阅读:15
作者:Derwood Pamphilon, Eileen Selogie, David McKenna, Jose A Cancelas-Peres, Zbigniew M Szczepiorkowski, Ron Sacher, John McMannis, Hermann Eichler, Henk Garritsen, Minoko Takanashi, Leo van de Watering, David Stroncek, Jo-Anna Reems

Aims

Wide acceptance of the colony-forming unit (CFU) assay as a reliable potency test for stem cell products is hindered by poor inter-laboratory reproducibility. The goal of this study was to ascertain current laboratory practices for performing the CFU assay with an eye towards identifying practices that could be standardized to improve overall reproducibility.

Conclusions

Survey results revealed inconsistent inter-laboratory practices for performing the CFU assay. The relatively low number of centers with validated CFU assays raises concerns about assay accuracy and emphasizes a need to establish central standards. The survey results shed light on numerous steps of the methodology that could be targeted for standardization across laboratories.

Methods

A survey to evaluate current laboratory practices for performing CFU assays was designed and internationally distributed.

Results

There were 105 respondents to the survey, of whom 68% performed CFU assays. Most survey recipients specified that an automated rather than a manual cell count was performed on pre-diluted aliquots of stem cell products. Viability testing methods employed various stains, and when multiple sites used the same viability stain, the methods differed. Cell phenotype used to prepare working cell suspensions for inoculating the CFU assay differed among sites. Most respondents scored CFU assays at 14-16 days of incubation, but culture plates were read with various microscopes. Of 57 respondents, 42% had not performed a validation study or established assay linearity. Only 63% of laboratories had criteria for determining if a plate was overgrown with colonies. Conclusions: Survey results revealed inconsistent inter-laboratory practices for performing the CFU assay. The relatively low number of centers with validated CFU assays raises concerns about assay accuracy and emphasizes a need to establish central standards. The survey results shed light on numerous steps of the methodology that could be targeted for standardization across laboratories.