Development of an In Vitro Model to Characterize the Effects of Transcatheter Aortic Valve on Coronary Artery Flow

建立体外模型以表征经导管主动脉瓣膜置换术对冠状动脉血流的影响

阅读:1

Abstract

Impairment of coronary artery flow, in either acute or chronic conditions, is a severe complication of transcatheter aortic valve (TAV) implantation, which can arise due to improper TAV positioning. However, little work has been done to quantify the effects of the TAV positioning on the coronary flow. In this study, a realistic in vitro model of coronary artery flow was developed and used to investigate the impact of TAV deployed orientations on coronary flow. The coronary hemodynamics was first replicated mathematically using a lumped parameter model with time-varying myocardial resistance. Based on the analytical model, two stepper motor controlled stopcock valves were integrated in a left heart simulator to represent the variable myocardial resistance in the experimental setup. The coronary flow and pressure waveforms obtained from the in vitro system were consistent with published data. With a TAV deployed in different orientations, the measured results demonstrated that TAV orientation does not have a significant impact on the coronary flow. The developed in vitro model can be further utilized to simulate coronary flow under various pathological conditions.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。