An ex vivo comparison of partial thromboplastin time and activated clotting time for heparin anticoagulation in an ovine model

在绵羊模型中,对肝素抗凝治疗的部分凝血活酶时间和活化凝血时间进行体外比较

阅读:1

Abstract

BACKGROUND: Sheep are a primary model of mechanical circulatory support (MCS) with heparin anticoagulation therapy frequently being monitored by activated clotting time (ACT) due to ease and cost. In patients undergoing long-term heparin therapy, other anticoagulation monitoring strategies, such as activated partial thromboplastin time (aPTT), have proven to be more reliable indicators for the adequacy of anticoagulation, frequently determined by heparin concentration. As there is a paucity of similar studies in sheep, we sought to investigate the correlation between heparin concentration and ACT and aPTT using whole sheep blood in an ex vivo model. METHODS: Fresh whole blood was serially drawn from an adult female Dorset-hybrid sheep and aliquots were placed into tubes containing heparin saline solutions with concentrations ranging from 0 to 7.81 U heparin per mL of whole blood. ACT and aPTT values were measured on each of the samples. The experiment was performed four times with the same animal. A simple linear regression was performed to determine correlation, and subgroup analysis was performed on low versus high heparin concentrations typically seen in human patients on long-term MCS, such as extracorporeal membrane oxygenation (ECMO), versus cardiopulmonary bypass, respectively. RESULTS: aPTT measurements versus the heparin concentration had an R(2)  = 0.7295. ACT measurements versus the heparin concentration had a R(2)  = 0.4628. aPTT measurements versus the ACT measurements had a R(2)  = 0.2974. The strength of the correlation between aPTT and heparin concentration increased at low heparin concentrations (R(2) = 0.8392). CONCLUSION: aPTT had a more reliable correlation to heparin concentration and thus anticoagulation level than ACT. This was particularly true at lower heparin concentrations, similar to ranges seen for patients on ECMO. The correlation between aPTT and ACT values was poor. Further in vivo studies should be performed to confirm our results.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。