Poor correlation between alcohol concentration in oral fluid and breath in subjects consuming beverages immediately before testing

受试者在测试前立即饮用饮料后,口腔液和呼气中的酒精浓度相关性较差

阅读:2

Abstract

INTRODUCTION: In previous research projects and clinical settings, alcohol analysis in oral fluid (saliva) has been used as an alternative to breath or blood alcohol testing. In this study we examined whether it is possible to obtain clinically relevant data regarding alcohol consumption in individuals who recently consumed alcohol by analysing oral fluid samples when the recommended rinsing of the mouth is impossible before sample collection. MATERIALS AND METHODS: We conducted a study of 89 nightclub patrons in Norway. Before collecting oral fluid samples and performing breath alcohol testing, participants were required to drink a glass of water to remove residual alcohol from the mouth. Oral fluid samples were collected with the Quantisal oral fluid collection device and analysed using an enzymatic method for alcohol. The alcohol concentration in the neat (undiluted) oral fluid was then calculated. Breath alcohol testing was performed using Lion Alcolmeter 500 instruments. RESULTS: No false-negative or false-positive results for alcohol were detected in the oral fluid when compared with those in the breath. The Intraclass Correlation Coefficient of 0.40 indicated a poor correlation between alcohol concentrations in the two sample types. CONCLUSIONS: The procedure for collecting oral fluid was suitable for the qualitative determination of alcohol intake but not for quantitative assessment. We recommend that oral fluid samples should not be used for estimating blood or breath alcohol concentrations in people who have recently consumed alcohol or non-alcoholic beverages, as recommended in the instructions for use.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。