Flushing of an intravenous catheter: a cause for unreliable laboratory results

冲洗静脉导管:导致实验室结果不可靠的原因之一

阅读:1

Abstract

INTRODUCTION: Phlebotomy is an error-prone process in which mistakes are difficult to reveal. This case report describes the effect on laboratory results originating from a blood sample collected in close proximity to an intravenous catheter. MATERIALS AND METHODS: A 69-year-old male patient was referred to the Emergency department where pneumonia was suspected. Phlebotomy was performed to collect blood samples to assess electrolytes, renal function, liver function, infection and haematological parameters. RESULTS: The laboratory analysis showed reduced potassium and calcium concentrations. To prevent life-threatening cardiac failure the clinician decided to correct those electrolytes. Remarkably, the electrocardiogram showed no abnormalities corresponding to hypokalaemia and hypocalcaemia. This observation, in combination with an overall increase in laboratory parameters with the exception of sodium and chloride, led to the suspicion of a preanalytical error. Retrospectively, an intravenous catheter was inserted in close proximity of the puncture place but no continuous infusion was started prior to phlebotomy. However, the intravenous catheter was flushed with sodium chloride. Since potential other causes were excluded, the flushing of the intravenous catheter with sodium chloride prior to phlebotomy was the most probable cause for the deviating laboratory results and subsequently for the unnecessary potassium and calcium suppletion. CONCLUSION: This case underlines the importance of caution in the interpretation of laboratory results obtained from specimens that are collected in the proximity of an intravenous catheter, even in the absence of continuous infusion.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。