Haemolysis and lipemia interfere with resistin and myeloperoxidase BioVendor ELISA assays

溶血和脂血会干扰抵抗素和髓过氧化物酶的BioVendor ELISA检测。

阅读:1

Abstract

INTRODUCTION: The aim of our study was to investigate the influence of haemolysis and lipemia on resistin (RES) and myeloperoxidase (MPO) measurement by BioVendor enzyme-linked immunosorbent assays (ELISA). MATERIALS AND METHODS: Blood was taken from healthy volunteers into lithium heparin tubes. Plasma samples were spiked with Lipofundin® emulsion (B. Braun Melsungen AG, Germany) for lipemia interference testing. Haemolysed samples were obtained by drawing aliquots of heparinized blood through a 26 gauge needle. Index of haemolysis (H), lipemia (L) and triglyceride concentration were measured on Abbott Architect c8000. Haemoglobin concentration was measured on Sysmex XN-1000. Concentrations of RES and MPO in all samples were determined with RES and MPO ELISA kits (BioVendor, Czech Republic). All measurements were performed in triplicate. Biases from the native samples were calculated for both analytes and compared with an arbitrary value (e.g. ± 10%). RESULTS: Triglyceride concentration in the investigated samples ranged from 0.57 to 38.23 mmol/L, which corresponds to L index from - 0.01 to 13.77. Haemoglobin concentration in all samples ranged from 0 to 8 g/L which correspond to H index from 0.05 to 8.77. Both MPO and RES showed significant biases at 1 g/L haemoglobin (58.7% and 66.7%, respectively). Also, both MPO and RES showed significant biases at 4.66 mmol/L triglycerides (33.8% and - 12.2%, respectively). CONCLUSIONS: Resistin BioVendor assays are affected by haemolysis and lipemia already at low degree of interferent. Haemolysis was found to interfere at 1 g/L haemoglobin for both assays, while lipemia interferes at 4.66 mmol/L of triglycerides.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。