Uncertainty Evaluation for the Quantification of Urinary Amphetamine and 4-Hydroxyamphetamine Using Liquid Chromatography-Tandem Mass Spectrometry: Comparison of the Guide to the Expression of Uncertainty in Measurement Approach and the Monte Carlo Method with R

利用液相色谱-串联质谱法定量分析尿液中苯丙胺和4-羟基苯丙胺的不确定度评估:测量不确定度表示指南方法与蒙特卡罗方法的比较(基于R²)。

阅读:1

Abstract

Estimating the measurement uncertainty (MU) is becoming increasingly mandatory in analytical toxicology. This study evaluates the uncertainty in the quantitative determination of urinary amphetamine (AP) and 4-hydroxyamphetamine (4HA) using a liquid chromatography-tandem mass spectrometry (LC-MS/MS) method based on the dilute-and-shoot approach. Urine sample dilution, preparation of calibrators, calibration curve, and method repeatability were identified as the sources of uncertainty. To evaluate the MU, the Guide to the Expression of Uncertainty in Measurement (GUM) approach and the Monte Carlo method (MCM) were compared using the R programming language. The MCM afforded a smaller coverage interval for both AP (94.83, 104.74) and 4HA (10.52, 12.14) than that produced by the GUM (AP (92.06, 107.41) and 4HA (10.21, 12.45)). The GUM approach offers an underestimated coverage interval for Type A evaluation, whereas the MCM provides an exact coverage interval under an abnormal probability distribution of the measurand. The MCM is useful in complex settings where the measurand is combined with numerous distributions because it is generated from the uncertainties of input quantities based on the propagation of the distribution. Therefore, the MCM is more practical than the GUM for evaluating the MU of urinary AP and 4HA concentrations using LC-MS/MS.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。