The Impact of Heli-FX EndoAnchor Application on Endograft Material: An Experimental Study

Heli-FX EndoAnchor应用对血管内移植物材料的影响:一项实验研究

阅读:1

Abstract

OBJECTIVE: The physical impact of the application of Heli-FX EndoAnchors (EA; Medtronic, Minneapolis, USA) on endograft (EG) material is unclear. This study aimed to examine the possible EG membrane alterations after EA implantation. METHODS: Heli-FX EndoAnchors were applied in vitro into four aortic endocuffs: AFX2 (Endologix Inc., Irvine, USA); Endurant II (Medtronic, Minneapolis, USA); Gore Excluder (W.L. Gore and Assoc., Flagstaff, USA); and Zenith Renu (Cook Aortic Interventions, Bloomington, USA). Two of these, Endurant II and Renu, are made of polyethylene terephthalate (PET), while Excluder and AFX2 are made of expanded polytetrafluoroethylene (ePTFE). The penetration angle was measured for each EA. The EAs were then carefully removed, and perforations examined with digital and fluorescent microscopy. The area and perimeter of the holes were digitally calculated, and material alterations were analysed. RESULTS: Of the 13 EAs applied, 12 remained in place. The mean penetration angle was 79°. The ePTFE perforations had oval openings, while PET perforations were round. After EA removal, comparisons between ePTFE and PET material perforations suggested a larger hole area (p = 0.011) and perimeter (p = 0.003) in the former. The ePTFE perforations in the AFX2 were the largest compared with the holes in other endocuffs (p = 0.050). The perforation channel of the ePTFE membrane of the Excluder cuff retained its form after EA removal. Local dissection like layer damage extended further. The perforations in both the Endurant II and the Renu endocuffs shared similar characteristics, with multiple fibres of PET elongated, distorted, or ruptured. CONCLUSION: During EA placement, the EG membrane undergoes local alteration and or destruction. Expanded PTFE, particularly AFX2 endocuffs (for which EA use is not recommended), are characterised by a more extensive degree of material alteration compared with PET. Additional studies are required to chronologically supplement these findings in fatigue tests.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。