Interfacial Strength Between Porous Metal Acetabular Components and Augments

多孔金属髋臼组件与增生物之间的界面强度

阅读:4

Abstract

BACKGROUND: The use of porous metal acetabular components combined with augments is a potential approach for revision hip arthroplasty in cases of severe bone defects. However, the strength of the augment-acetabular cup interface has not been studied previously. This study evaluated the interfacial strength between acetabular cups and augments. METHODS: The strength of the interface between the augment, bone cement, and the acetabular cup was tested using lever-out test. The effects of contact area and type of bone cement on the interface strength were investigated. RESULTS: The interfaces exhibited excellent stability. With the same cup-augment combination and different bone cements with different properties, no statistical differences were observed in main mechanical indicators (p > 0.05); however, when the same bone cement was used with different cup-augment combinations, the maximum moment and yield moment decreased significantly with a decrease in the contact area (p < 0.05). The interface stiffness exhibited a decreasing trend, although the difference was not statistically significant. Failure always occurred on the cement-acetabular cup side. CONCLUSION: The interface between the acetabular cup and augment is stable and primarily influenced by the size of the contact area, not the type and mechanical properties of bone cement. CLINICAL RELEVANCE: This study provides a theoretical basis for the use of augments during revision surgery. The factors affecting the strength were explored. This could promote the improvement of prostheses and guide the design of a new generation of prostheses.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。