Anesthetic gas consumption with target-controlled administration versus a semi-closed circle system with automatic end-tidal concentration control in an artificial lung model

在人工肺模型中,采用靶控给药与采用自动呼气末浓度控制的半封闭循环系统进行麻醉气体消耗量的比较

阅读:1

Abstract

The use of volatile anesthetics as sedatives in the intensive care unit is relevant to the patient's outcome. We compared anesthetic gas consumption of the conventional semi-closed Aisys CS(TM) with the MIRUS(TM) system, which is the first anesthetic gas reflector system that can administer desflurane in addition to isoflurane and sevoflurane. We connected an artificial lung model to either a MIRUS(TM) system and a Puritan Bennett(TM) 840 ventilator or an Aisys CS(TM) anesthesia machine. We found that consumption of 0.5% isoflurane, which corresponds to the target concentration 0.5 MAC, was averaged to 2 mL/h in the MIRUS(TM) system, which is identical to the Aisys CS(TM) at a fresh gas flow (FGF) of 1.0 L/min. MIRUS(TM) consumption of 1% sevoflurane was averaged to 10 mL/h, which corresponds to 8.4 mL/h at FGF 2.5 L/min. The MIRUS(TM) system consumed 3% or 4% desflurane at an average of 13.0 mL/h or 21.3 mL/h, which is between the consumption at 1.0 L/min and 2.5 L/min FGF. Thus, the MIRUS(TM) system can effectively deliver volatile anesthetics in clinically relevant concentrations in a similar rate as a conventional circular breathing system at FGFs between 1.0 L/min and 2.5 L/min.