Development and evaluation of time-resolved fluorescent immunochromatographic assay for quantitative detection of SARS-CoV-2 spike antigen

时间分辨荧光免疫层析法定量检测 SARS-CoV-2 刺突抗原的开发与评价

阅读:8
作者:Buzhou Xu, Hao Tang, Yiming Weng, Valerie Sloane Jones, Shuhong Luo, Chih Yun Cho, Yongping Lin, Jianmin Fang, Xuedong Song, Ruo-Pan Huang

Background

The spread of COVID-19 worldwide caused by the severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) has necessitated efficient, sensitive diagnostic

Conclusion

The new S1-RBD antigen test is a rapid (15 min), sensitive, and specific assay that requires minimal sample preparation. Critically, the assay correlated closely with PCR-based methodology in nasopharyngeal swab samples, showing that the detected S1-RBD antigen levels correlate with SARS-CoV-2 virus load. Therefore, the new TRF lateral flow test for S1-RBD has potential application in point-of-care settings.

Methods

We constructed and evaluated a portable, disposable lateral flow device, which detected the S1-RBD protein directly in nasopharyngeal swab samples. The device emits a fluorescent signal in the presence of S1-RBD, which can be captured by an automated TRF instrument.

Results

The TRF lateral flow assay signal was linear from 0 to 20 ng/ml and demonstrated high accuracy and reproducibility. When evaluated with clinical nasopharyngeal swabs, the assay was performed at >80% sensitivity, >84% specificity, and > 82% accuracy for detection of the S1-RBD antigen.