Better safe than sorry? Results from an ex-vivo study demonstrate that the thulium fiber laser may cause eye injury without standard protection

防患于未然总是好的?一项离体研究结果表明,如果没有标准防护措施,铥光纤激光可能会造成眼损伤。

阅读:2

Abstract

INTRODUCTION: We conducted a study using an ex-vivo porcine model to evaluate whether a thulium fiber laser (TFL) induces ocular injury in the context of inadvertent exposure to the laser beam. METHODS: A 365 μm TFL was positioned at a set distance (0 cm, 5 cm, 8 cm, and 10 cm) from a freshly harvested (<12 hours) porcine eyeball and the laser was activated for one second at select laser settings for lithotripsy (0.2 J at 50 Hz, 0.5 J at 20 Hz, and 1 J at 10 Hz) and soft tissue ablation (2 J at 10 Hz, 1 J at 50 Hz). The experiment was repeated with laser safety goggles and prescription eyeglasses. Thermal injury was assessed by histopathological analysis. RESULTS: Without eye protection, corneal injury was observed even at 10 cm away for one lithotripsy setting (1 J at 10 Hz) and both tissue ablation settings. All thermal injuries observed were superficial only, except for at 0 cm distance, where deep-layer injury was observed. Laser safety goggles offered complete protection regardless of setting or distance. Partial protection was demonstrated with prescription glasses: histopathological damage was observed for both soft tissue ablation settings and only at 0 cm for two lithotripsy settings (0.5 J at 20 Hz, 1 J at 10 Hz). CONCLUSIONS: The TFL can induce ocular injury at close distances and at higher power settings. The use of laser safety goggles confers complete protection while prescription eyeglasses confer partial protection. Further study is warranted.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。