Handheld Echocardiography in a Clinical Practice Scenario: Concordances Compared to Standard Echocardiographic Reports

手持式超声心动图在临床实践中的应用:与标准超声心动图报告的一致性比较

阅读:3

Abstract

BACKGROUND: The purpose of this study was to assess the utility of a handheld device (HH) used during common daily practice and its agreement with the results of a standard echocardiography study (STD) performed by experienced sonographers and echocardiographer. METHODS: A prospective follow-up was conducted in an adult outpatient echocardiography clinic. Experienced sonographers performed the STD and an experienced echocardiographer performed the HH. STD included 2-dimensional images, Doppler and hemodynamics analysis. Hemodynamic assessment was not performed with the HH device because the HH does not include such technology. The images were interpreted by blinded echocardiographers, and the agreement between the reports was analyzed. RESULTS: A total of 108 patients were included; and the concordance for left ventricle (LV) ejection fraction (EF), wall motion score index, LV and right ventricle (RV) function, RV size, and mitral and aortic stenosis was excellent with κ values greater than 0.80. Wall motion abnormalities had good concordance (κ value 0.78). The agreement for LV hypertrophy, mitral and aortic regurgitation was moderate, and tricuspid and pulmonary regurgitation agreements were low (κ values of 0.26 and 0.25, respectively). CONCLUSIONS: In a daily practice scenario with experienced hands, HH demonstrated good correlation for most echocardiography indications, such as ventricular size and function assessment and stenosis valve lesion analyses.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。