Noninvasive measurement of (13)Carbon turnover for evaluation of porcine renal grafts during ex vivo machine perfusion

无创测量(13)碳周转率以评估离体机械灌注过程中猪肾移植的效果

阅读:1

Abstract

BACKGROUND: Kidney transplantation suffers from a shortage of donor organs. Despite this, a lot of grafts are discarded due to inadequate quality. As many kidneys are afflicted by transient filtration failure early after preservation, classical renal function tests are not applicable to differentiate between prospective recovery or continuing deficit of renal function. METHODS: Using normothermic machine perfusion as a platform for pre-implantation evaluation of the graft, we present a novel evaluative approach based on the metabolic turnover of (13)C-acetate during isolated perfusion. After injection of the tracer, (13)CO(2) as a metabolic end-product can be quantified by high-precision laser-based spectroscopy in the gas outflow of the oxygenator. Three groups of porcine kidneys with graduated ischemic injury were investigated. RESULTS: This quantitative approach is able to discriminate acceptable quality kidneys, most likely to recover within days from poor kidney grafts that are unlikely to regain notable glomerular function with high discriminatory power (area under the ROC curve 0.91; P < 0.001 By contrast, conventional renal function tests are rather ineffective under these circumstances. CONCLUSIONS: This assessment method offers the potential to quantitatively assess donor kidney quality using a measurable output, salvaging donors that would otherwise have been discarded.