Sweat Cortisol and Cortisone Determination in Healthy Adults: UHPLC-MS/MS Assay Validation and Clinical Application

健康成人汗液皮质醇和可的松的测定:UHPLC-MS/MS检测方法的验证及临床应用

阅读:1

Abstract

A simple and effective ultra-high-performance liquid chromatography assay linked to tandem mass spectrometry (UHPLC-MS/MS) for measuring cortisol and cortisone levels in human sweat has been developed and validated. A noninvasive world standard sweat collecting equipment was utilized to collect samples. The samples were analyzed using an Atlantis dC18 (2.1 × 100 mm, 3 μm) column with a 2 mM ammonium acetate and acetonitrile (1 : 1, v : v) mobile phase. In an isocratic condition, the mobile phase was delivered at a flow rate of 0.3 ml/minute. A positive electrospray ionization interface with multiple-reaction monitoring mode was used to provide simultaneous quantification of cortisol, cortisone, and internal standard at transitions of 363.11 to 121.00, 361.18 to 163.11, and 367.19 to 121.24, respectively. The method was validated for cortisol and cortisone determination over a concentration range of 0.5-50 ng/mL The detection limits for cortisol and cortisone in human sweat were 0.3 and 0.2 ng/ml, respectively. The interday coefficients of variation of cortisol and cortisone were ≤8.5% and ≤10.01%, whereas bias was in the range from -7.9% to 2.1% and from -4.3% to 3.0%, respectively. The assay was successfully applied to evaluate the cortisol-to-cortisone ratio in sweat samples collected from healthy adult volunteers.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。