A 0.85% saline as alternative detection buffer for SD-Bioline HIV rapid test in resource limited setting

在资源有限的环境下,0.85%生理盐水可作为SD-Bioline HIV快速检测的替代检测缓冲液。

阅读:1

Abstract

Accuracy in the diagnosis is a key step to identify HIV infected individuals for appropriate management. Insufficient supply of manufacturer´s buffers in relation to the number of strips per kit has negative impact on patient´s results hence improper patient´s management. In resource limited settings, some laboratory staff use different substitute buffers which has never been validated on their reliability. This study aimed at comparing the performance of 0.85% saline and SD-Bioline manufacturer´s buffer in detection of HIV antibodies. A total of 220 whole blood specimens: 110 HIV positive specimens from patients attending care and treatment center (CTC) and 110 HIV negative specimens from blood donors were re-tested for HIV status using SD-Bioline HIV rapid test using manufacturer´s buffer and 0.85% saline separately. Data and laboratory results were recorded in Microsoft excel sheet followed by analysis using STATA version 13. For all tested samples, the level of agreement between 0.85% saline and manufacturer´s buffer was 98.64% (kappa=0.9727). The value of kappa indicates very good agreement between 0.85% saline and manufacturer´s buffer. In incidents where manufacturer´s buffer is not sufficient, 0.85% saline can give reliable results. Further studies to evaluate the suitable buffer for other rapid tests for HIV and other diseases are recommended especially in resource limited settings.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。