Performance Evaluation of Four Deep Learning-Based CAD Systems and Manual Reading for Pulmonary Nodules Detection, Volume Measurement, and Lung-RADS Classification Under Varying Radiation Doses and Reconstruction Methods

在不同辐射剂量和重建方法下,对四种基于深度学习的计算机辅助诊断系统和人工判读方法在肺结节检测、体积测量和肺部影像报告和数据系统(Lung-RADS)分类方面的性能进行评估

阅读:1

Abstract

Background: Optimization of pulmonary nodule detection across varied imaging protocols remains challenging. We evaluated four DL-CAD systems and manual reading with volume rendering (VR) for performance under varying radiation doses and reconstruction methods. VR refers to a post-processing technique that generates 3D images by assigning opacity and color to CT voxels based on Hounsfield units. Methods: An anthropomorphic phantom with 169 artificial nodules was scanned at three dose levels using two kernels and three reconstruction algorithms (1080 image sets). Performance metrics included sensitivity, specificity, volume error (AVE), and Lung-RADS classification accuracy. Results: DL-CAD systems demonstrated high sensitivity across dose levels and reconstruction settings, with three fully automatic DL-CAD systems (0.92-0.95) outperforming manual CT readings (0.72), particularly for sub-centimeter nodules. However, DL-CAD systems exhibited limitations in volume measurement and Lung-RADS classification accuracy, especially for part-solid nodules. VR-enhanced manual reading outperformed original CT interpretation in nodule detection, particularly benefiting less-experienced radiologists under suboptimal imaging conditions. Conclusions: These findings underscore the potential of DL-CAD for lung cancer screening and the clinical value of VR in low-dose settings, but they highlight the need for improved classification algorithms.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。