A novel multimodality anthropomorphic phantom enhances compliance with quality assurance guidelines for magnetic resonance imaging in radiotherapy

一种新型多模态人体模型提高了放射治疗磁共振成像质量保证指南的依从性。

阅读:1

Abstract

BACKGROUND AND PURPOSE: The use of magnetic resonance imaging (MRI) for radiotherapy (RT) simulation has grown, prompting quality assurance (QA) guidelines by the Institute of Physics and Engineering in Medicine (IPEM) and the American Association of Physicists in Medicine (AAPM). This study compares a novel multimodality anthropomorphic phantom to an American College of Radiology (ACR) phantom for a subset of these MRI-specific QA tests in RT. MATERIALS AND METHODS: A novel 3D-printed multimodality head-and-neck anthropomorphic phantom was compared to an ACR large MRI phantom. IPEM and AAPM-recommended QA tests were conducted, including informatics/connectivity/data transfer, MRI-CT registration, end-to-end QA, and signal-to-noise ratio (SNR)/percentage integral uniformity (PIU) assessments using RT accessories. RESULTS: Both phantoms were suitable for informatics/connectivity/data transfer. In MRI-CT registration, no errors were found; the ACR phantom offered more quantitative landmarks, while the anthropomorphic phantom provided limited structures. Both phantoms achieved target registration errors (TREs) below 0.97 mm and dice similarity coefficient (DSC) values above 0.9, meeting guidelines. For end-to-end QA, the anthropomorphic phantom facilitated dose measurements of 1.994 Gy versus a calculated 2.01 Gy (-0.8 %). SNR and PIU assessments showed higher values in radiology setups compared to RT setups for both phantoms. CONCLUSIONS: Multimodality anthropomorphic phantoms compatible with dosimetric equipment allow realistic end-to-end QA, unlike the ACR phantom. While the ACR phantom is suitable for informatics and MRI-CT registration, anthropomorphic phantoms better represent clinical scenarios. For comprehensive QA, both ACR and anthropomorphic phantoms are required. Additionally, large field-of-view (FOV) phantoms are crucial for evaluating large FOV MRI distortions.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。