In vitro measurements of ultrafiltration precision in hemofiltration and hemodialysis devices used in infants, Part 2: Comparison of PrisMax and CARPEDIEM with previous data on NIDUS, Prismaflex and Aquarius

体外测量用于婴儿的血液滤过和血液透析装置的超滤精度,第二部分:PrisMax 和 CARPEDIEM 与先前关于 NIDUS、Prismaflex 和 Aquarius 的数据比较

阅读:1

Abstract

BACKGROUND: We sought to determine in vitro whether the PrisMax and CARPEDIEM hemofiltration and hemodialysis devices can reliably deliver ultrafiltration (UF) control that is sufficiently precise to treat infants. METHODS: We have previously measured the precision of UF control of the Prismaflex, Aquarius and NIDUS devices by in vitro testing with a bag of saline set up as a dummy patient, and comparing the differences between the UF set and displayed by the devices, and the actual fluid removal or addition measured by precise weighing. Here we have tested the PrisMax (updated version of Prismaflex) and the CARPEDIEM using the same method. RESULTS: The variances of the setting vs. actual errors, and display vs. actual errors after 15 min of 'treatment' with the PrisMax and CARPEDIEM were similar, but were significantly larger than in the NIDUS, and much smaller than in the Prismaflex. However, after a 4-h 'treatment session', the cumulative errors were still within ± 9 mL for these devices, compared with a maximum error of 2.6 mL in the NIDUS, and a deviation of -37.5 mL in the Prismaflex. CONCLUSIONS: The PrisMax and the CARPEDIEM have adequate precision to be used in infants. The only device with UF error below 3 ml in 4 h is the volumetrically-controlled NIDUS. We recommend that regulatory bodies should introduce UF precision-testing for devices intended for use in infants.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。