Evaluation of intra-fractional CBCT image quality using different megavoltage beam sizes and imaging conditions in a phantom study

在体模研究中,使用不同兆伏级光束尺寸和成像条件评估分次内锥形束CT图像质量

阅读:1

Abstract

The intra-fractional cone beam computed tomography can provide real-time anatomical information while saving treatment time, leading to important breakthroughs in image-guided radiation therapy. However, MV-level photon scattering during the image acquisition will disrupt the image quality. We obtained intra-fractional cone beam computed tomography images of a standard phantom using different sizes of MV-beam and imaging conditions such as imaging time, tube current and tube voltage. We analyzed the uniformity index, noise power spectrum, low contrast visibility and spatial resolution of the images to investigate the effect of MV-level photon scattering on the image quality of intra-fractional cone beam computed tomography. It was found that these parameters of the images varied regularly with the size of the MV-beam, which can help us to predict in advance whether the quality of the images meets the clinical requirements based on the radiation treatment plan to decide whether to use intra-fractional cone beam computed tomography. The research also shows that the different imaging conditions can improve or deteriorate the image parameter, which can also help us choose reasonable imaging conditions to obtain better image quality according to the clinical situation.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。