Worsening Heart Failure due to AV Dissociation From Safety Mode in CRT-P: A Case Requiring Urgent Generator Replacement

CRT-P 患者因房室分离脱离安全模式导致心力衰竭加重:一例需紧急更换发生器的病例

阅读:1

Abstract

Boston Scientific reported that dual chamber INGENIO family pacemakers or cardiac resynchronization therapy pacemakers (CRT-P) may initiate safety mode. Once Safety Mode is activated, telemetry with the programmer becomes unavailable. This case highlights worsening heart failure due to loss of atrioventricular synchronization associated with transition to Safety Mode and VVI pacing. Urgent admission for generator replacement was necessary because mode adjustment was not possible.