Abstract
Heart rate variability (HRV) is a key physiological marker for autonomic nervous system function and cardiovascular health. Photoplethysmography (PPG) is commonly used to derive HRV metrics in wearable and low-cost monitoring systems. This study presents a comparative assessment of basic HRV metrics obtained from a MAX30102 optical sensor and a custom analog circuitry with an ADS1115 analog-to-digital converter (ADC). Both measurement pathways were carefully aligned using analog high-pass and low-pass filters and a consistent digital filtering pipeline, ensuring that the frequency bands relevant to HRV were preserved. PPG signals were recorded simultaneously, and inter-beat intervals were extracted to calculate the Standard Deviation of NN intervals (SDNN), Root Mean Square of Successive Differences (RMSSD), and Percentage of successive NN intervals >50 ms (pNN50) across multiple 30-s windows. Bland–Altman analysis was employed to evaluate agreement between the two methods. Results indicate that the analog circuit with an ADS1115 achieves comparable HRV basic metrics to the MAX30102 sensor, with improved Signal-to-Noise Ratio (SNR) due to high-resolution ADC and low-noise analog amplification. These findings demonstrate that a carefully designed analog acquisition system can reliably reproduce HRV basic parameters from PPG signals, providing an alternative approach for low-cost, flexible biosensing platforms.