Accuracy and Reproducibility of Handheld 3D Ultrasound Versus Conventional 2D Ultrasound for Urinary Bladder Volume Measurement: A Prospective Comparative Study

手持式三维超声与传统二维超声测量膀胱容量的准确性和可重复性:一项前瞻性比较研究

阅读:1

Abstract

Background/Objectives: Accurate urinary bladder (UB) volume measurement is essential for diagnosing urinary retention, evaluating post-void residuals, and guiding catheterization decisions. Conventional 2D ultrasound and automated non-visual bladder scanners can be limited by operator variability and systematic errors. Three-dimensional (3D) ultrasound may improve accuracy and reproducibility, but data on handheld, semi-automated devices remain scarce. This study aimed to compare the accuracy, reproducibility, and feasibility of a handheld 3D ultrasound device versus conventional 2D ultrasound for UB volume estimation, using measured voided volume as the reference standard. Methods: Fifty-three healthy male volunteers (mean age 19.6 ± 2.0 years) underwent bladder volume assessment by two novice operators using both methods: 2D ultrasound (manual caliper-based) and handheld 3D ultrasound device (Butterfly iQ). Each operator performed two repeated measurements per method. True voided volume was recorded immediately after scanning. Accuracy was assessed using median differences, percentage error, and Bland-Altman analysis. Intra- and inter-operator reproducibility were evaluated with intraclass correlation coefficients (ICC). Results: Both methods significantly underestimated bladder volume (p < 0.001). The 3D method demonstrated higher accuracy, with a median percentage error of -11.2% to -12.0%, versus -27.6% to -36.7% for 2D. The mean bias ranged from -64.9 mL to -72.3 mL for 3D, compared to -137.4 mL to -191.6 mL for 2D. Intra-operator reproducibility was excellent for all methods (ICC > 0.96). Inter-operator agreement was higher for 3D (ICC = 0.977; bias 7.3 mL) than for 2D (ICC = 0.927; bias -54.2 mL). All scans were completed successfully; however, the 3D device occasionally faced technical errors in large bladder volumes. Conclusions: Handheld 3D ultrasound yielded greater accuracy and inter-operator consistency than conventional 2D ultrasound in healthy adults, with minimal operator input. Both methods underestimated true volume, indicating the need for clinical consideration when interpreting measurements. These findings support broader clinical adoption of handheld 3D ultrasound, particularly in settings with variable sonographic expertise, while highlighting the need for validation in elderly and pathological populations.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。