Comparative Evaluation of Classic Mechanical and Digital Goldmann Applanation Tonometers

经典机械式和数字式戈尔德曼压平式眼压计的比较评价

阅读:3

Abstract

Objectives: The objective of this study was to evaluate the agreement and clinical interchangeability of intraocular pressure (IOP) measurements obtained with the mechanical Haag-Streit AT900 Goldmann applanation tonometer (mGAT) and the digital Huvitz HT5000 applanation tonometer (dGAT). Methods: This retrospective comparative study included 53 eyes of 28 patients undergoing routine ophthalmologic evaluation. Each eye underwent IOP measurement using both mGAT and dGAT in a randomized sequence. Central corneal thickness (CCT) was also recorded. Pearson's correlation coefficient was used to determine correlation between paired IOP measurements. Bland-Altman plots were graphed for the analysis of differences for IOP between the instruments. Results: A total of 53 eyes of 28 patients (15 males) were included in the study. The mean age of the patients was 62.6 years. The mean mGAT and dGAT measurements were 16.3 ± 6.6 mmHg (range 9-50) and 16.4 ± 6.2 mmHg (range 8.8-45.9), respectively (p = 0.53). A strong, significant positive correlation was found for paired IOP measurements by the two instruments (r = 0.98; p < 0.0001). Bland-Altman analysis revealed 95% limits of agreement from -2.5 to +2.3 mmHg, with a small but statistically significant proportional bias favoring mGAT at higher IOP levels. Additionally, 91% of paired measurements were within ±2 mmHg. CCT-related differences were statistically and clinically insignificant. Conclusions: IOP measurements obtained with mGAT and dGAT were highly correlated and clinically interchangeable for the range tested. The Huvitz HT5000 may serve as a reliable alternative to the classic Goldmann tonometer in routine clinical settings.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。