Clinical evaluation of the installation conditions and accuracy of wearable neonatal jaundice meter

对可穿戴式新生儿黄疸仪的安装条件和准确性进行临床评估

阅读:1

Abstract

Neonatal jaundice is a physiological phenomenon that occurs in newborns. However, because excess bilirubin can cause cytotoxicity, abnormal hyperbilirubinemia can lead to bilirubin encephalopathy. In the evaluation of neonatal jaundice, serum bilirubin concentrations are indirectly assessed via a transcutaneous bilirubinometer. However, commercially available transcutaneous bilirubinometers are expensive, especially in developing countries. Therefore, many newborns still suffer from bilirubin encephalopathy worldwide. We previously developed a prototype wearable jaundice meter. In this study, we aimed to investigate further improvements in this device and evaluate its measurement accuracy at different measurement sites and at different pressures. The bilirubin values of forty-eight newborns were measured at different sites (middle of the forehead and upper eyebrows) and with different pressures (1 and 2 newtons). The transcutaneous bilirubin values measured by the wearable bilirubin meter were strongly correlated with the transcutaneous bilirubin values measured by a commercially available bilirubin meter (r > 0.9) and the actual serum bilirubin values (> 0.84), even under different conditions. This wearable jaundice meter is much cheaper and smaller than commercially available jaundice meters and is a novel medical device that could be useful for evaluating neonatal jaundice.