Evaluation of a Breath-Indicating Device for Enhanced Respiratory Monitoring and Apnea Detection in Postoperative Care: A Comparative Study

评估一种用于增强术后护理呼吸监测和呼吸暂停检测的呼吸指示装置:一项比较研究

阅读:1

Abstract

OBJECTIVES: Accurate respiratory monitoring is crucial in post-anesthetic care settings due to increased risks of respiratory complications. This study evaluates the impact of a new breath-indicating device, ApnoLight (PEMDx Pty Ltd, Brisbane, Queensland, Australia), on medical staff's accuracy in recording respiratory rates and detecting apneic events, along with the device's acceptance among nurses. METHODS: Twenty-five nurses from a hospital in Brisbane, Australia, participated. A simulated patient was fitted with the ApnoLight device on an oxygen mask. Nurses conducted six respiratory rate observations at varying distances (bedside, two meters, and five meters), both with and without the device. The patient's respiratory rates varied from eight to 25 breaths per minute. The accuracy of respiratory rate recordings and the time to identify apnea events were compared between simple observation and device-assisted observations. RESULTS: The ApnoLight significantly reduced error rates in respiratory rate recordings: 27.58% at the bedside (P = 0.09, t = 1.31), 90.99% at two meters (P = 0.02, t = 1.98), and 96.37% at five meters (P = 0.0002, t = 4.02). The mean time to identify apnea decreased from 12.96 ± 9.12 seconds (simple observation) to 7.42 ± 2.19 seconds (with ApnoLight device). All apnea events were identified with the device, whereas four were undetected without it. Feedback showed that 96% (N = 24) of nurses found the device improved respiratory rate accuracy, and 100% (N = 25) found it made apnea identification easier. CONCLUSIONS: The ApnoLight device has the potential to enhance respiratory rate monitoring accuracy and apnea detection in postoperative settings. Its implementation could improve patient safety and streamline clinical workflows.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。