Determination of irinotecan, SN-38 and SN-38 glucuronide using HPLC/MS/MS: Application in a clinical pharmacokinetic and personalized medicine in colorectal cancer patients

使用 HPLC/MS/MS 测定伊立替康、SN-38 和 SN-38 葡萄糖醛酸:在结直肠癌患者的临床药代动力学和个性化医疗中的应用

阅读:20
作者:Chalirmporn Atasilp, Pichai Chansriwong, Ekapob Sirachainan, Thanyanan Reungwetwattana, Apichaya Puangpetch, Santirhat Prommas, Suwannee Sirilerttrakul, Budsaba Rerkarmnuaychoke, Sansanee Wongwaisayawan, Chonlaphat Sukasem

Background

Irinotecan (CPT-11) is chemotherapy used mainly in the metastatic colorectal cancer. The

Conclusion

This method was successfully used to quantify CPT-11, SN-38, and SN-38G and applied to a pharmacokinetic study.

Methods

A 100 μL of plasma was prepared after protein precipitation and analyzed on a C18 column using 0.1% acetic acid in water and 0.1% acetic acid in acetonitrile as mobile phases. The mass spectrometer worked with multiple reaction monitoring (MRM) in positive scan mode. The standard curves were linear on a concentration range of 5-10 000 ng/mL for CPT-11, 5-1000 ng/mL for SN-38, and 8-1000 ng/mL for SN-38G.

Results

In this assay, the intra and interday precision consisted of ≤9.11% and ≤11.29% for CPT-11, ≤8.70% and 8.31% for SN-38, and ≤9.90 and 9.64% for SN-38G.