Validity and Agreement of a Portable Dynamometer in Evaluating Hip Strength and Rate of Torque Development: Influence of Fixation Methods

便携式测力计在评估髋关节力量和扭矩发展速率方面的有效性和一致性:固定方式的影响

阅读:2

Abstract

BACKGROUND: Portable dynamometers (PD) are widely used to assess hip muscle strength; however, few studies have explored their validity and agreement when evaluating the rate of torque development (RTD) of hip muscles, particularly with different fixation methods. HYPOTHESIS/PURPOSE: The purpose of this study was to determine the validity and agreement of isometric hip strength and RTD using the PD with two fixation methods: placing the PD on the body (PDpush) or a treatment table (PDpull), compared to an isokinetic dynamometer (IKD). It was hypothesized that an PD would be a valid tool for assessing hip strength and RTD, with PDpull demonstrating strong validity and PDpush showing moderate validity; and its assessments would yield higher values than those obtained from IKD. STUDY DESIGN: Cross-sectional study. METHODS: Twenty healthy, physically active individuals (10 females and 10 males) were included in this study. Hip abduction and extension strength (peak torque) and RTD (early phase: RTD0-100 ms and late phase: RTD100-200 ms) were assessed using PDpush, PDpull, and an IKD. RESULTS: Moderate-to-strong correlations were found between PD (both PDpush and PDpull) and IKD for all hip strength (r=0.46-0.73; p<0.05) and RTD (r=0.45-0.73, ρ=0.80; p<0.05) measures. PDpull demonstrated strong correlations for hip abduction RTD0-100 ms and hip extension strength and RTD100-200 ms (r=0.70-0.73; p<0.001), whereas PDpush showed moderate correlations (r=0.45-0.58; p<0.05). Bland-Altman plots with mean differences ± 95% limits of agreement indicated that the PDpush method overestimated all hip abduction and extension strength and RTD as compared to IKD. The PDpull method overestimated hip abduction and extension strength but had mixed results for RTD, either underestimating hip abduction RTD while overestimating hip extension RTD. CONCLUSION: The PD is valid tool for evaluating hip strength and RTD, with PDpull offering strong validity for certain measures; however, PD measurements did not agree with the absolute values obtained from IKD. LEVEL OF EVIDENCE: Level 3.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。