Mixed gain detector configurations for time-resolved X-ray solution scattering

用于时间分辨X射线溶液散射的混合增益探测器配置

阅读:2

Abstract

X-ray detection at X-ray free-electron lasers is challenging in part due to the XFEL's extremely short and intense X-ray pulses. Experimental measurements are further complicated by the large fluctuations inherent to the self-amplified spontaneous emission process producing the X-rays. At the Linac Coherent Light Source the ePix10ka2M detector offers multiple gain modes, and auto-ranging between these, to increase the dynamic range while retaining low noise. For diffuse scattering techniques, such as time-resolved X-ray solution scattering, where the shape of the scattering pattern largely does not change between exposures, a fixed mix of different gain modes offers many of the same advantages as auto-ranging. We find that configuring individual ASICs in separate gain modes does not impact the intensity linearity of the gain response and has a limited effect on the effective dynamic range in regions with different gain mode settings while avoiding the complexities of auto-ranging. Small (<5%) non-linear gain contributions arise when pixels on the same ASIC are configured in different gain modes. We present a configuration scheme that is designed to select the optimal mixed gain configuration to minimize effects of saturation in the high-/medium-gain region, while maximizing the number of pixels with higher gain to improve the signal-to-noise ratio.