Alterations in implantable cardioverter defibrillator lead parameters following left ventricular assist device implantation

左心室辅助装置植入后植入式心脏复律除颤器导线参数的变化

阅读:2

Abstract

BACKGROUND: Left ventricular assist devices (LVADs) are increasingly used in the management of advanced heart failure. The majority of these patients have pre-existing implantable cardioverter defibrillators (ICDs). The proximity between the LVAD inflow cannula and right ventricular (RV) defibrillation lead raises the potential for disruption of ICD function. METHODS: This is a retrospective analysis of 95 patients with ICDs at a single tertiary care center who underwent LVAD implantation and who met inclusion criteria. The primary outcome was changes in the pre-operative and post-operative transvenous ICD RV lead parameters. These changes were stratified by the age of the RV lead and analyzed via a paired t-test. The secondary outcome was disruption to the ICD requiring an intervention. RESULTS: LVAD implantation was associated with significant decreases in sensed amplitude (p < 0.01) and high voltage impedance (p < 0.01) and an increase in capture threshold (p = 0.017). When stratified by age of the RV lead, patients with leads older than two years had similar trends in all parameters. However, RV leads that were two years old or younger only showed a significant change in high voltage impedance (p < 0.01). Mechanical disruption of the ICD related to the surgery was infrequent but significant. CONCLUSION: Because LVAD implantation is capable of impacting ICD function and causing mechanical disruption, close monitoring should be paid to the ICD in the peri-operative period including obtaining a full interrogation.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。