[Spherical measurement-based analysis of gradient nonlinearity in magnetic resonance imaging]

[基于球面测量的磁共振成像梯度非线性分析]

阅读:1

Abstract

The gradient field, one of the core magnetic fields in magnetic resonance imaging (MRI) systems, is generated by gradient coils and plays a critical role in spatial encoding and the generation of echo signals. The uniformity or linearity of the gradient field directly impacts the quality and distortion level of MRI images. However, traditional point measurement methods lack accuracy in assessing the linearity of gradient fields, making it difficult to provide effective parameters for image distortion correction. This paper introduced a spherical measurement-based method that involved measuring the magnetic field distribution on a sphere, followed by detailed magnetic field calculations and linearity analysis. This study, applied to assess the nonlinearity of asymmetric head gradient coils, demonstrated more comprehensive and precise results compared to point measurement methods. This advancement not only strengthens the scientific basis for the design of gradient coils but also provides more reliable parameters and methods for the accurate correction of MRI image distortions.