Occlusion of the Ascenda Catheter in a Patient with Pump Twiddler's Sydrome: A Case Report

泵扭动综合征患者Ascenda导管阻塞:病例报告

阅读:1

Abstract

INTRODUCTION: Intrathecal baclofen (ITB) therapy is an effective way to manage spasticity in numerous conditions, including multiple sclerosis, stroke, and cerebral palsy. While pump failure is a common complication of ITB, improvements in device design have led to reduction of complications. In particular, the Ascenda catheter from Medtronic, Inc. was designed to resist kinking and associated complications; indeed, no incidences of catheter twisting or occlusion have been reported in literature prior to this case. CASE REPORT: We report a case of a 32-year old gentleman who presented to the clinic with symptoms of baclofen withdrawal 19 months after he had a programmable pump implanted for spasticity. During the diagnostic evaluation it was discovered that the patients pump had flipped in his abdominal pocket. He was taken to surgery to reorient the pump, during which time it was noted the catheter was tightly coiled on itself occluding flow. The twisted catheter was excised and replaced with a new segment.His symptoms subsequently resolved. CONCLUSIONS: Although catheter occlusions have subsided since the approval of the Ascenda catheter, pump twiddler's syndrome remains a risk factor for this complication. This is the first report describing this syndrome in a patient with the Ascenda catheter.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。