Enhanced-Precision Measurement of Glutathionyl Hemoglobin by MALDI-ToF MS

利用MALDI-ToF MS提高谷胱甘肽血红蛋白的测量精度

阅读:1

Abstract

Glutathionyl-hemoglobin (HbSSG) is used as a human biomarker to pinpoint systemic oxidative stress caused by various pathological conditions, noxious lifestyles, and exposure to drugs and environmental or workplace toxicants. Measurement by MALDI mass spectrometry is most frequently used, however, the method suffers from excessive uncontrolled variability. This article describes the improvement of a MALDI-ToF mass spectrometry method for HbSSG measurement through enhanced precision, based on strict control of sample preparation steps and spreadsheet-based data analysis. This improved method displays enhanced precision in the analysis of several hundred samples deriving from studies in different classes of healthy and diseased human subjects. Levels span from 0.5% (lower limit of detection) up to 30%, measured with a precision (as SE%) < 0.5%. We optimized this global procedure to improve data quality and to enable the Operator to work with a reduced physical and psychological strain. Application of this method, for which full instruction and the data analysis spreadsheet are supplied, can encourage the exploitation of HbSSG to study human oxidative stress in a variety of pathological and living conditions and to rationally test the efficacy of antioxidant measures and treatments in the frame of health promotion.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。