Transcatheter para mitral annular ring leak device closure - A case report

经导管二尖瓣环旁漏封堵术——病例报告

阅读:1

Abstract

Para mitral annular ring leakage can occur following ring dehiscence after mitral annuloplasty. Percutaneous device closure of para-annular ring leakage can be performed successfully to treat such regurgitations with good transesophageal echocardiography guidance and patient selection. While para valvular device closure has been described in the medical literature, there have been few anecdotal reports published on para ring leak device closures. In this case, we highlight our experience from the successful closure of a para mitral annular ring closure with an AVP III device. The patient had a para annular ring regurgitation post coronary artery bypass grafting with mitral ring annuloplasty presenting with hemolytic anemia and acute renal failure, successfully treated by percutaneous device closure. LEARNING OBJECTIVE: This report describes the safety and effectiveness of a transcatheter para ring leak closure with an AVP III device. We applied the principles of device closure of paravalvular leak from our experience and related data from literature for this case and describe various hardware and techniques used for a successful closure of a para mitral ring leak.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。