The Association Between Ventilator Dyssynchrony, Delivered Tidal Volume, and Sedation Using a Novel Automated Ventilator Dyssynchrony Detection Algorithm

利用新型自动化呼吸机不同步检测算法研究呼吸机不同步、潮气量输送和镇静之间的关联

阅读:1

Abstract

OBJECTIVE: Ventilator dyssynchrony is potentially harmful to patients with or at risk for the acute respiratory distress syndrome. Automated detection of ventilator dyssynchrony from ventilator waveforms has been difficult. It is unclear if certain types of ventilator dyssynchrony deliver large tidal volumes and whether levels of sedation alter the frequency of ventilator dyssynchrony. DESIGN: A prospective observational study. SETTING: A university medical ICU. PATIENTS: Patients with or at risk for acute respiratory distress syndrome. INTERVENTIONS: Continuous pressure-time, flow-time, and volume-time data were directly obtained from the ventilator. The level of sedation and the use of neuromuscular blockade was extracted from the medical record. Machine learning algorithms that incorporate clinical insight were developed and trained to detect four previously described and clinically relevant forms of ventilator dyssynchrony. The association between normalized tidal volume and ventilator dyssynchrony and the association between sedation and the frequency of ventilator dyssynchrony were determined. MEASUREMENTS AND MAIN RESULTS: A total of 4.26 million breaths were recorded from 62 ventilated patients. Our algorithm detected three types of ventilator dyssynchrony with an area under the receiver operator curve of greater than 0.89. Ventilator dyssynchrony occurred in 34.4% (95% CI, 34.41-34.49%) of breaths. When compared with synchronous breaths, double-triggered and flow-limited breaths were more likely to deliver tidal volumes greater than 10 mL/kg (40% and 11% compared with 0.2%; p < 0.001 for both comparisons). Deep sedation reduced but did not eliminate the frequency of all ventilator dyssynchrony breaths (p < 0.05). Ventilator dyssynchrony was eliminated with neuromuscular blockade (p < 0.001). CONCLUSION: We developed a computerized algorithm that accurately detects three types of ventilator dyssynchrony. Double-triggered and flow-limited breaths are associated with the frequent delivery of tidal volumes of greater than 10 mL/kg. Although ventilator dyssynchrony is reduced by deep sedation, potentially deleterious tidal volumes may still be delivered. However, neuromuscular blockade effectively eliminates ventilator dyssynchrony.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。