Microstructure Optimization for Design of Porous Tantalum Scaffolds Based on Mechanical Properties and Permeability

基于力学性能和渗透性的多孔钽支架微观结构优化设计

阅读:1

Abstract

Porous tantalum (Ta) implants have important clinical application prospects due to their appropriate elastic modulus, and their excellent bone growth and bone conduction ability. However, porous Ta microstructure designs generally mimic titanium (Ti) implants commonly used in the clinic, and there is a lack of research on the influence of the microstructure on the mechanical properties and penetration characteristics, which will greatly affect bone integration performance. This study explored the effects of different microstructure parameters, including the fillet radius of the middle plane and top planes, on the mechanics and permeability properties of porous Ta diamond cells through simulation, and put forward an optimization design with a 0.5 mm midplane fillet radius and 0.3 mm top-plane fillet radius in order to significantly decrease the stress concentration effect and improve permeability. On this basis, the porous Ta structures were prepared by Laser Powder Bed Fusion (LPBF) technology and evaluated before and after microstructural optimization. The elastic modulus and the yield strength were increased by 2.31% and 10.39%, respectively. At the same time, the permeability of the optimized structure was also increased by 8.25%. The optimized microstructure design of porous Ta has important medical application value.