A Nonlinear Three-Dimensional Finite Element Analysis of Stress Distribution and Microstrain Evaluation in Short Dental Implants with Three Different Implant-Abutment Connections in Single and Splinted Conditions in the Posterior Mandible

后牙区下颌骨三种不同种植体-基台连接方式的短种植体单颗及夹板连接状态下应力分布及微应变的非线性三维有限元分析

阅读:1

Abstract

BACKGROUND: Stress distribution plays a vital role in the longevity and success of implant-supported prosthesis. This study evaluated the von Mises stress and microstrain in the peri-implant bone and the implant-abutment junction of short dental implants with three different implant-abutment connections in splinted and unsplinted conditions using finite element analysis (FEA). MATERIALS AND METHODS: In this experimental study, nine transversely isotropic finite element models were developed, and randomly divided into three equal groups (n = 3): control, (Group AC) single-standard 4.3 × 10 mm bone level implant-supported restorations with external hexagonal (EH) connection, internal conical (IC) and internal trichannel (ITC) connection, single short implant-supported restorations (Group AT), and splinted short implant-supported restorations (Group B) for each of the three implant-abutment connections, respectively. A 200 N load was applied along the long axis of the implants and a 100 N (45°) oblique load was applied and von Mises stress and microstrain values were evaluated. RESULTS: Single standard implants demonstrated the highest von Mises stress and microstrain values followed by single short implants and splinted short implants, respectively. Among the implant-abutment connections, the IC connection showed the highest values and the ITC connection showed the least values. CONCLUSION: Within the limitations of this study, it was concluded that splinting of short dental implants demonstrated lesser and more homogeneous stress and microstrain, especially on oblique loading. The microstrain values for all connections evaluated were within the physiological loading limit (200-2,500 N) and were hence considered safe for clinical use.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。