Automated synthesis of [68Ga]Ga-FAPI-46 without pre-purification of the generator eluate on three common synthesis modules and two generator types

使用三种常见的合成模块和两种发生器类型自动合成 [68Ga]Ga-FAPI-46,无需对发生器洗脱液进行预纯化

阅读:3
作者:Ammar Alfteimi #, Ulf Lützen #, Alexander Helm, Michael Jüptner, Marlies Marx, Yi Zhao, Maaz Zuhayra

Background

The recent development of quinoline-based radiotracers, which act as fibroblast activation protein inhibitors (FAPIs), has shown promising preclinical and clinical advantages. [68Ga]Ga-FAPI-46 is a new radiotracer for in vivo detection of the fibroblast activation protein by positron emission tomography (PET). Recently, the automated synthesis of [68Ga]Ga-FAPI-46 was reported based on pre-concentration and purification of the generator eluate by using a cation exchange-cartridge. Our

Conclusions

Independent of the synthesis modules or the generators employed we were able to complete the synthesis of [68Ga]Ga-FAPI-46 in 12 min including the process of purification and formulation. In all cases, the final products showed more than 99.5% chemical purity and the radiochemical yield reached around 92.5% (decay corrected). All quality control parameters (e.g. sterility, stability and radiochemical purity) were conform to the European Pharmacopoeia.

Results

We developed and evaluated the GMP compliant automatic synthesis of [68Ga]Ga-FAPI-46 using two different 68Ge/68Ga generators (an Eckert & Ziegler, GalliaPharm generator, 1.85 GBq/50 mCi and an iThemba generator, 1.85 GBq/50 mCi) Somerset West, South Africa) and three different commercial and customized systems: the EasyOne module from Trasis; the GaSy module from Synthra with a customized synthesis template and a customized single use cassette. Additionally, the automatic synthesis of [68Ga]Ga-FAPI-46 was established on a GallElut synthesis module from Scintomics with fixed tubing. Conclusions: Independent of the synthesis modules or the generators employed we were able to complete the synthesis of [68Ga]Ga-FAPI-46 in 12 min including the process of purification and formulation. In all cases, the final products showed more than 99.5% chemical purity and the radiochemical yield reached around 92.5% (decay corrected). All quality control parameters (e.g. sterility, stability and radiochemical purity) were conform to the European Pharmacopoeia.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。