Simulation, Fabrication and Analysis of Silver Based Ascending Sinusoidal Microchannel (ASMC) for Implant of Varicose Veins

用于静脉曲张植入物的银基上升正弦微通道 (ASMC) 的模拟、制造和分析

阅读:10
作者:Muhammad Javaid Afzal, Shahzadi Tayyaba, Muhammad Waseem Ashraf, M Khalid Hossain, M Jalal Uddin, Nitin Afzulpurkar

Abstract

Bioengineered veins can benefit humans needing bypass surgery, dialysis, and now, in the treatment of varicose veins. The implant of this vein in varicose veins has significant advantages over the conventional treatment methods. Deep vein thrombosis (DVT), vein patch repair, pulmonary embolus, and tissue-damaging problems can be solved with this implant. Here, the authors have proposed biomedical microdevices as an alternative for varicose veins. MATLAB and ANSYS Fluent have been used for simulations of blood flow for bioengineered veins. The silver based microchannel has been fabricated by using a micromachining process. The dimensions of the silver substrates are 51 mm, 25 mm, and 1.1 mm, in length, width, and depth respectively. The dimensions of microchannels grooved in the substrates are 0.9 mm in width and depth. The boundary conditions for pressure and velocity were considered, from 1.0 kPa to 1.50 kPa, and 0.02 m/s to 0.07 m/s, respectively. These are the actual values of pressure and velocity in varicose veins. The flow rate of 5.843 (0.1 nL/s) and velocity of 5.843 cm/s were determined at Reynolds number 164.88 in experimental testing. The graphs and results from simulations and experiments are in close agreement. These microchannels can be inserted into varicose veins as a replacement to maintain the excellent blood flow in human legs.

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。