DEMONSTRATION AND EXPERIMENTAL VALIDATION OF PLASTIC-ENCASED RESONANT ULTRASONIC PIEZOELECTRIC ACTUATOR FOR MRI-GUIDED SURGICAL ROBOTS

用于MRI引导手术机器人的塑料封装谐振超声压电致动器的演示和实验验证

阅读:1

Abstract

Intra-operative medical imaging based on magnetic resonance imaging (MRI) coupled with robotic manipulation of surgical instruments enables precise feedback-driven procedures. Electrically powered non-ferromagnetic motors based on piezoelectric elements have shown to be well suited for MRI robots. However, even avoiding ferrous materials, the high metal content on commercially available motors still cause distortions to the magnetic fields. We construct semi-custom piezoelectric actuators wherein the quantity of conductive material is minimized and demonstrate that the distortion issues can be partly addressed through substituting several of these components for plastic equivalents, while maintaining motor functionality. Distortion was measured by assessing the RMS change in position of 49 centroid points in a 12.5mm square grid of a gelatin-filled phantom. The metal motor caused a distortion of up to 4.91mm versus 0.55mm for the plastic motor. An additional SNR drop between motor off and motor spinning of approximately 20% was not statistically different for metal versus plastic (p=0.36).

特别声明

1、本页面内容包含部分的内容是基于公开信息的合理引用;引用内容仅为补充信息,不代表本站立场。

2、若认为本页面引用内容涉及侵权,请及时与本站联系,我们将第一时间处理。

3、其他媒体/个人如需使用本页面原创内容,需注明“来源:[生知库]”并获得授权;使用引用内容的,需自行联系原作者获得许可。

4、投稿及合作请联系:info@biocloudy.com。